Advertisement

Syntax
Parameter
Query response
Example
SCPI Commands
:PAGE:MEASure:PGUSetup:PULSe:PGU<n>:PEAK
:PAGE:MEASure:PGUSetup:PULSe:PGU<n>:PEAK
This command sets the PEAK VALUE of PGU<n>.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
You use this command only if the specified PGU is set to VPULse mode (not V) by
the :PAGE:CHAN:PGU<n>:MODE command.
At *RST, this value is 0.1 V.
:PAGE:MEASure:PGUSetup:PULSe:PGU<n>:PEAK peak | MINimum |
MAXimum
Parameter
Type
peak
numeric
peak <newline><^END>
peak is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:PEAK 5"
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:PEAK MAX"
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:PEAK?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:PULS:PGU1:PEAK? MAX"
ENTER @Hp4155;A
5-184
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
-40 to 40 V