Syntax
Parameter
Query response
Example
SCPI Commands

:PAGE:CHANnels:UFUNction:DEFine

:PAGE:CHANnels:UFUNction:DEFine
This command defines the USER FUNCTION.
If the specified function name already exists, this command overwrites the old data
with the newly defined data.
If the specified function name does not exist, the new name appears in the
uppermost blank field on the CHANNELS:USER FUNCTION DEFINITION page.
A maximum of six user functions can be defined. If this limit is exceeded,
Unable to define more than 6 User functions.
At *RST, all defined functions are deleted.
:PAGE:CHANnels:UFUNction:DEFine name, unit, definition
For query:
:PAGE:CHANnels:UFUNction:DEFine? name
Parameter
Type
name
string
unit
string
definition
string
unit, definition <newline><^END>
unit and definition are string response data, but do not contain double quote
characters at the beginning and end.
OUTPUT @Hp4155;":PAGE:CHAN:UFUN:DEF 'VTH','V','@L1X'"
OUTPUT @Hp4155;":PAGE:CHAN:UFUN:DEF? 'VTH'"
ENTER @Hp4155;A$
5-102
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
String of up to 6 alphanumeric characters.
1st character must be alphabet.
String of up to 6 alphanumeric characters.
The null string is allowed.
expression
error (-182) is generated.