Syntax
Parameter
Query response
Example

:DATA|:TRACe:UNIT

This command sets the unit for data of the specified user variable.
The user variable must first be defined by the :DATA | :TRACe:DEFine or
:PAGE:CHANnels:UVARiable:DEFine command, or on the CHANNELS: USER
VARIABLE DEFINITION page.
:DATA | :TRACe:UNIT user_variable_name ,unit
Parameter
user_variable_name
unit
unit <newline><^END>
OUTPUT @Hp4155;":TRAC:UNIT 'Vdata','mV'"
OUTPUT @Hp4155;":TRAC:UNIT? 'Vdata'"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Type
string or
user variable name to specify unit for
character
string
unit of the user variable data
SCPI Commands
:DATA|:TRACe:UNIT
Explanation
5-17