Syntax
Parameter
Query response
Example
SCPI Commands
:PAGE:MEASure[:SWEep]:VAR1:PCOMpliance:STATe
:PAGE:MEASure[:SWEep]:VAR1:PCOMpliance:STA
Te
This command sets the power compliance of VAR1 to disable.
If SMU for VAR1 unit is set to VPULse or IPULse mode, STATe is set to OFF. You
cannot set power compliance for VAR1.
If the unit type of VAR1 is VSU, STATe is ignored.
If power compliance value is set by PCOMpliance pcompliance, the query returns
1.
At *RST, this value is OFF.
:PAGE:MEASure [ :SWEep] :VAR1:PCOMpliance:STATe OFF | 0
Parameter
Type
OFF or 0
boolean
0 | 1 <newline><^END>
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:PCOM:STATE OFF"
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:PCOM:STATE?"
ENTER @Hp4155;A
5-228
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
power compliance is disabled