Syntax
Example
Syntax
Parameter
Query response
Example
:PAGE:DISPlay:ANALysis[:MENU]
This command changes the present display page to DISPLAY: ANALYSIS SETUP
page.
This command does not have query form.
:PAGE:DISPlay:ANALysis [ :MENU]
OUTPUT @Hp4155;":PAGE:DISP:ANAL"
:PAGE:DISPlay[:SETup]:DRESOL
This command changes the resolution of the measurement data displayed on the
screen. This command also has a query form.
At *RST, this value is set to set to NORM.
:PAGE:DISPlay [:SETup] :DRESOL NORMal | EXTend
:PAGE:DISPlay:SETup:DRESOL?
Parameter
Type
NORMal
character
EXTend
character
NORM | EXT <newline><^END>
OUTPUT @Hp4155;":PAGE:DISP:SET:DRESOL EXT"
OUTPUT @Hp4155;":PAGE:DISP:SET:DRESOL?"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:DISPlay:ANALysis[:MENU]
Explanation
The resolution depends on the measurement
resolution described in the specifications. For
example, 1 fA resolution on the 10 pA range.
The resolution is extended up to the full scale of
the internal A/D converter. For example, 10 aA
resolution on the 10 pA range.
SCPI Commands
5-121