Syntax
Parameter
Query response
Example
Syntax
Example
Syntax
SCPI Commands
:PAGE:MEASure:SAMPling:IINTerval
:PAGE:MEASure:SAMPling:IINTerval
This command sets the INITIAL INTERVAL for sampling measurement.
:PAGE:MEASure:SAMPling:IINTerval initial_interval | MINimum | MAXimum
Parameter
initial_interval
numeric
initial_interval <newline><^END>
initial_interval is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:IINT 0.5"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:IINT MIN"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:IINT?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:IINT? MIN"
ENTER @Hp4155;A
:PAGE:MEASure:SAMPling[:MENU]
This command changes the present display page to MEASURE: SAMPLING
SETUP page.
This command does not have query form.
:PAGE:MEASure:SAMPling [ :MENU]
OUTPUT @Hp4155;":PAGE:MEAS:SAMP"
:PAGE:MEASure:SAMPling:MODE
This command sets the MODE for sampling measurement.
The sampling mode determines the sampling interval.
:PAGE:MEASure:SAMPling:MODE LINear | L10 | L25 | L50 | THINnedout
5-212
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Type
60E-6 to 65.535 (s)
Explanation