Page:measure[:Sweep]:Sstop - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example
OUTPUT @Hp4155;":PAGE:MEAS:PULS:WIDTH MAX"
OUTPUT @Hp4155;":PAGE:MEAS:PULS:WIDTH?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:PULS:WIDTH? MAX"
ENTER @Hp4155;A

:PAGE:MEASure[:SWEep]:SSTop

This command sets the sweep stop mode for an abnormal status.
The abnormal statuses that can be detected are as follows:
SMU reaches its compliance setting.
Current of VSU exceeds ±100 mA.
SMU or VSU oscillates.
A/D converter overflow occurs.
Average current of PGU exceeds ±100 mA.
At *RST, this value is OFF.
If you set the SERIES RESISTANCE of SMU<n> to a non-zero value by
:PAGE:CHANnels[:CDEFinition]:SMU<n>:SRESistance, this value is
automatically changed to COMP.
:PAGE:MEASure [ :SWEep] :SSTop ABNormal | COMPliance | OFF
Parameter
Type
ABNormal
character
COMPliance
character
OFF
character
If you set the power compliance for VAR1, you cannot specify OFF.
OFF | ABN | COMP <newline><^END>
OUTPUT @Hp4155;":PAGE:MEAS:SST ABN"
OUTPUT @Hp4155;":PAGE:MEAS:SST?"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:MEASure[:SWEep]:SSTop
Explanation
Stop sweep if abnormal status is detected.
Stop sweep if some SMU reaches its compliance
setting.
Continue sweep even if abnormal status is detected.
SCPI Commands
5-225