Syntax
Parameter
Query response
Example
Syntax
Example
SCPI Commands
:PAGE:MEASure:PGUSetup:CONStant:PGU<n>[:SOURce]
:PAGE:MEASure:PGUSetup:CONStant:PGU<n>[:SO
URce]
This command sets the constant SOURCE value of PGU<n>.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
You use this command only if the specified PGU is set to V mode (not VPULSE) by
the :PAGE:CHAN:PGU<n>:MODE command.
At *RST, this value is 0.
:PAGE:MEASure:PGUSetup:CONStant:PGU<n> [ :SOURce] source | MINImum |
MAXimum
Parameter
Type
source
numeric
source <newline><^END>
source is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:CONS:PGU1 10"
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:CONS:PGU1 MAX"
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:CONS:PGU1?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:PGUS:CONS:PGU1? MAX"
ENTER @Hp4155;A
:PAGE:MEASure:PGUSetup[:MENU]
This command changes the present display page to MEASURE: PGU SETUP.
This command does not have query form.
:PAGE:MEASure:PGUSetup [ :MENU]
OUTPUT @Hp4155;":PAGE:MEAS:PGUS"
5-178
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
-40 to 40 V