Page:stress:setup:constant:pgu<N>[:Source] - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Example
Syntax
Parameter
Query response
Example
OUTPUT @Hp4155;":PAGE:STR:SET:ACC RES"
OUTPUT @Hp4155;":PAGE:STR:SET:ACC?"
ENTER @Hp4155;A
:PAGE:STRess:SETup:CONStant:PGU<n>[:SOURce]
This command sets the constant SOURCE value of PGU<n> for stress.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
You use this command only if the specified PGU is set to V mode (not VPULSE) by
the :PAGE:STR:PGU<n>:MODE command.
At *RST, this value is 0.
:PAGE:STRess:SETup:CONStant:PGU<n> [ :SOURce] source | MINimum |
MAXimum
Parameter
Type
source
numeric
source <newline><^END>
source is NR3 response data type.
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:PGU1 10"
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:PGU1 MAX"
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:PGU1?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:STR:SET:CONS:PGU1? MAX"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:STRess:SETup:CONStant:PGU<n>[:SOURce]
-40 to 40 V
SCPI Commands
Explanation
5-265