:PAGE:MEASure:SAMPling:FILTer

This command sets the SMU output filter state for sampling measurement.
:PAGE:MEASure:SAMPling:FILTer OFF | ON | 0 | 1
Parameter
Type
OFF or 0
boolean
ON or 1
boolean
0 | 1 <newline><^END>
OUTPUT @Hp4155;":PAGE:MEASure:SAMPling:FILT ON"
OUTPUT @Hp4155;":PAGE:MEASure:SAMPling:FILT?"
ENTER @Hp4155;A
:PAGE:MEASure:SAMPling:HTIMe
This command sets the HOLD TIME of sampling measurement.
:PAGE:MEASure:SAMPling:HTIMe hold_time | MINimum | MAXimum
Parameter
Type
hold_time
numeric
hold_time <newline><^END>
hold_time is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:HTIM 1.5"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:HTIM MAX"
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:HTIM?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:SAMP:HTIM? MAX"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:MEASure:SAMPling:FILTer
Explanation
Filter off
Filter on
Explanation
3E-2 to 655.35 (s). Resolution: 100 ms
SCPI Commands
5-211