Page:stress:setup:pulse:pgu<N>:Width - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Parameter
Query response
Example
Syntax
Parameter
Query response
Example
Parameter
Type
trailing_time
numeric
trailing_time <newline><^END>
trailing_time is NR3 response data type.
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:TRA 0.001"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:TRA MAX"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:TRA?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:TRA? MAX"
ENTER @Hp4155;A
:PAGE:STRess:SETup:PULSe:PGU<n>:WIDTh
This command sets the pulse WIDTH of PGU<n> for stress.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
The mode of the specified PGU must be VPULse.
At *RST, this value is 5 ms.
:PAGE:STRess:SETup:PULSe:PGU<n>:WIDTh width | MINimum | MAXimum
Parameter
Type
width
numeric
width <newline><^END>
width is NR3 response data type.
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:WIDTH 0.005"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:WIDTH MIN"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:WIDTH?"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:STRess:SETup:PULSe:PGU<n>:WIDTh
Explanation
100 ns to 10 ms
Explanation
1E-6 to 9.99 s
SCPI Commands
5-277

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4156c

Table of Contents