:PAGE:CHANnels[:CDEFinition]:PGU<n>:VNAMe
:PAGE:CHANnels[:CDEFinition]:PGU<n>:VNAMe
This command sets the VNAME of PGU<n>.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
At *RST, this value is not defined.
:PAGE:CHANnels [ :CDEFinition] :PGU<n>:VNAMe name
Parameter
Type
name
string
name <newline><^END>
name is string response data, but does not contain double quote characters at the
beginning and end of the string.
OUTPUT @Hp4155;":PAGE:CHAN:PGU1:VNAME 'VD'"
OUTPUT @Hp4155;":PAGE:CHAN:PGU1:VNAME?"
ENTER @Hp4155;A$
:PAGE:CHANnels[:CDEFinition]:SMU<n>:DISable
This command deletes the settings of SMU<n>.
<n> is required to specify SMU number. Valid SMU numbers are SMU1 through
SMU6, depending on the configuration.
This command does not have query form.
:PAGE:CHANnels [ :CDEFinition] :SMU<n>:DISable
OUTPUT @Hp4155;":PAGE:CHAN:SMU1:DIS"
5-88
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
String of up to 6 alphanumeric characters.
1 st character must be alphabet.