Table 3-2
Standard Event Status Register
Bit
Definition
0
Operation Complete
(OPC)
1
Request Control
(RQC)
2
Query Error (QYE)
3
Device Dependent
Error (DDE)
4
Execution Error
(EXE)
5
Command Error
(CME)
6
Not Used
7
Power On (PON)
8 to 15
Reserved
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Status Reporting Structure
Standard Event Status Register
Explanation
This event bit has meaning only if a request to
monitor is set by the *OPC command. Refer to
"*OPC" earlier in this chapter. This bit is set to
1 if there are no pending operations.
This event bit indicates to the controller that
instrument is requesting permission to become
the active controller-in-charge.
•
An attempt is being made to read data
from the Output Queue when no data is
present or pending.
•
Data in the Output Queue has been lost.
This event bit indicates that an error has
occurred which is neither a Command Error, a
Query Error, or an Execution Error.
Syntax of command is correct, but cannot be
executed due to some condition of instrument.
A command syntax error has been detected.
Always 0.
This event bit indicates that an off-to-on
transition has occurred in instrument's power
supply.
Always 0.