Syntax
Query response
Example
Syntax
Query response
Example
SCPI Commands

:PAGE:MEASure:MSETup:ITIMe:MEDium:APERture?

:PAGE:MEASure:MSETup:ITIMe:MEDium:APERtur
e?
This command returns the MEDIUM integration time in terms of the APERTURE,
which means time.
This command has query form only.
At *RST, this value is 1/selected line frequency.
:PAGE:MEASure:MSETup:ITIMe:MEDium:APERture?
aperture <newline><^END>
aperture is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:MED:APER?"
ENTER @Hp4155;A
:PAGE:MEASure:MSETup:ITIMe:MEDium:NPLCycl
es?
This command returns the MEDIUM integration time in terms of the Number of
Power Line Cycles.
The returned value is always 1 because the medium table is fixed.
This command has query form only.
At *RST, this value is 1.
:PAGE:MEASure:MSETup:ITIMe:MEDium:NPLCycles?
nplcycles <newline><^END>
nplcycles is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:MED:NPLC?"
ENTER @Hp4155;A