Syntax
Parameter
Query response
Example
Syntax
Parameter
Query response

:PAGE:KSWeep:CURSor:X

This command positions the cursor at the specified X coordinate. This command
does not affect the Y coordinate of the cursor.
At *RST, the cursor is located at the center of the graph.
:PAGE:KSWeep:CURSor:X x_position | MINimum | MAXimum
Parameter
Type
x_position
numeric
x_position <newline><^END>
x_position is NR3 response data.
OUTPUT @Hp4155;":PAGE:KSW:CURS:X 1.35"
OUTPUT @Hp4155;":PAGE:KSW:CURS:X MAX"
OUTPUT @Hp4155;":PAGE:KSW:CURS:X?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:KSW:CURS:X? MIN"
ENTER @Hp4155;A
:PAGE:KSWeep:CURSor:Y
This command positions the cursor at the specified Y coordinate. This command
does not affect the X coordinate of the cursor.
At *RST, the cursor is located at the center of the graph.
:PAGE:KSWeep:CURSor:Y y_position | MINimum | MAXimum
Parameter
Type
y_position
numeric
y_position <newline><^END>
y_position is NR3 response data type.
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:KSWeep:CURSor:X
Explanation
desired X coordinate of cursor
Explanation
desired Y coordinate of cursor