Page:channels[:Cdefinition]:Vmu<N>:Disable; Page:channels[:Cdefinition]:Vmu<N>:Mode - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Example
Syntax
Parameter
Query response
Example
:PAGE:CHANnels[:CDEFinition]:VMU<n>:DISable
This command deletes the settings of VMU<n>.
<n> is required to specify VMU number. Valid VMU numbers are VMU1 through
VMU2.
This command does not have query form.
:PAGE:CHANnels [ :CDEFinition] :VMU<n>:DISable
OUTPUT @Hp4155;":PAGE:CHAN:VMU1:DIS"
:PAGE:CHANnels[:CDEFinition]:VMU<n>:MODE
This command sets the MODE of VMU<n>.
<n> is required to specify VMU number. Valid VMU numbers are VMU1 through
VMU2.
At *RST, this value is V.
:PAGE:CHANnels [ :CDEFinition] :VMU<n>:MODE V | DVOLt
Parameter
Type
V
character
DVOLt
character
V | DVOL | DIS <newline><^END>
If VMU is not used, the response data may be DIS(DISable).
OUTPUT @Hp4155;":PAGE:CHAN:VMU1:MODE V"
OUTPUT @Hp4155;":PAGE:CHAN:VMU1:MODE?"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:CHANnels[:CDEFinition]:VMU<n>:DISable
Explanation
voltage measurement mode
differential voltage measurement mode
SCPI Commands
5-95