Syntax
Query response
Example
Syntax
Example
:PAGE:MEASure:MSETup:ITIMe:SHORt:NPLCycles
?
This command returns the SHORT integration time in terms of the Number of
Power Line Cycles.
The NPLCycle value is related to APERture by the equation:
NPLCycles = APERture ´ selected line frequency
This command has query form only.
At *RST, this value is 640 ms ´ selected line frequency.

:PAGE:MEASure:MSETup:ITIMe:SHORt:NPLCycles?

nplcycle <newline><^END>
nplcycle is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:SHORT:NPLC?"
ENTER @Hp4155;A
:PAGE:MEASure:MSETup[:MENU]
This command changes the present display page to MEASURE: MEASURE
SETUP page.
This command does not have query form.
:PAGE:MEASure:MSETup [ :MENU]
OUTPUT @Hp4155;":PAGE:MEAS:MSET"
Agilent 4155C/4156C SCPI Command Reference, Edition 1