Page:ksweep:var1:Points - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example

:PAGE:KSWeep:VAR1:POINts

This command sets the number of steps for VAR1 of knob sweep measurement.
At *RST, this value is 101.
:PAGE:KSWeep:VAR1:POINts points | MINimum | MAXimum
Parameter
Type
points
numeric
points <newline><^END>
points is NR1 response data type.
OUTPUT @Hp4155;":PAGE:KSW:VAR1:POIN 101"
OUTPUT @Hp4155;":PAGE:KSW:VAR1:POIN MAX"
OUTPUT @Hp4155;":PAGE:KSW:VAR1:POIN?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:KSW:VAR1:POIN?MAX"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:KSWeep:VAR1:POINts
Explanation
1 to 1001