Syntax
Parameter
Query response
Example
SCPI Commands

:PAGE:MEASure:QSCV:LCANcel

:PAGE:MEASure:QSCV:LCANcel
This command enables or disables leakage current compensation for Quasi-static
CV measurements. This command also has a query form.
At *RST, this value is OFF.
:PAGE:MEASure:QSCV:LCANcel OFF | ON | 0 | 1
:PAGE:MEASure:QSCV:LCANcel?
Parameter
Type
OFF or 0
Boolean
ON or 1
Boolean
0 | 1 <newline><^END>
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:LCAN ON"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:LCAN?"
ENTER @Hp4155;A
5-196
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
Leakage current compensation is disabled.
Leakage current compensation is enabled.