Page:measure:qscv:range - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example
SCPI Commands

:PAGE:MEASure:QSCV:RANGe

:PAGE:MEASure:QSCV:RANGe
This command sets the measurement RANGE of the measurement channel used in
the Quasi-static CV measurements. This command also has a query form.
At *RST, this value is set to 1E-9.
:PAGE:MEASure:QSCV:RANGe range | MINimum | MAXimum
:PAGE:MEASure:QSCV:RANGe?
Parameter
Type
range
numeric
range <newline><^END>
range is an NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:RANG MAX"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:RANG?"
ENTER @Hp4155;A
5-198
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
10E-12, 100E-12, 1E-9, or 10E-9 for the 4156C.
1E-9 or 10E-9 for the 4155C, MPSMU, HPSMU.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4156c

Table of Contents