Page:measure:msetup:zcancel[:State] - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Query response
Example
See also
Syntax
Parameter
Query response
Example
0 | 1 <newline><^END>
Parameter
Type
0
boolean
1
boolean
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ZCAN:SMU1?"
ENTER @Hp4155;A
:PAGE:MEAS:MSET:ZCAN[:STATe]

:PAGE:MEASure:MSETup:ZCANcel[:STATe]

This command controls whether the ZERO OFFSET CANCEL mode is enabled.
At *RST, this value is OFF.
:PAGE:MEASure:MSETup:ZCANcel [ :STATe] OFF | ON | 0 | 1
Parameter
Type
OFF or 0
boolean
ON or 1
boolean
0 | 1 <newline><^END>
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ZCAN ON"
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ZCAN?"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:MEASure:MSETup:ZCANcel[:STATe]
Explanation
zero offset cancel is not enabled.
zero offset cancel is enabled.
Explanation
zero offset cancel is not enabled for any units.
zero offset cancel is enabled depending on the
settings of each unit.
SCPI Commands
5-173