:PAGE:MEASure[:SWEep]:VAR2:STEP

This command sets the sweep STEP value of VAR2.
You use this command only if there is an SMU or VSU whose function (FCTN) is
VAR2.
At *RST, this value is 20 mA.
:PAGE:MEASure [ :SWEep] :VAR2:STEP step | MINimum | MAXimum
Parameter
Type
step
numeric
step <newline><^END>
step is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:STEP 20E-6"
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:STEP?"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:MEASure[:SWEep]:VAR2:STEP
Explanation
-400 to 400 V or -2 to 2 A. The range of this value
depends on the unit type of VAR2.