:PAGE:MEASure:QSCV:CONStant:SMU<n>:COMPliance
:PAGE:MEASure:QSCV:CONStant:SMU<n>:COMPliance
This command sets the COMPLIANCE value of the source SMU<n> for the
Quasi-static CV measurement. This command also has a query form.
<n> is required to specify the SMU number. Valid SMU numbers are SMU1 through
SMU6, depending on the configuration.
Use this command only if the function of the specified SMU is CONStant and the
mode is not COMMon.
:PAGE:MEASure:QSCV:CONStant:SMU<n>:COMPliance compliance |
MINimum | MAXimum
:PAGE:MEASure:QSCV:CONStant:SMU<n>:COMPliance?
Parameter
Type
compliance
numeric
compliance <newline><^END>
compliance is an NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:CONS:SMU1:COMP 0.1"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:CONS:SMU1:COMP?"
ENTER @Hp4155;A
5-190
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
–200 to 200 V or –1 to 1 A. The range of this
value depends on the SMU type being used.