Syntax
Parameter
Query response
Example
:PAGE:MEASure[:SWEep]:CONStant:SMU<n>:COMPliance
:PAGE:MEASure[:SWEep]:CONStant:SMU<n>:COM
Pliance
This command sets the constant COMPLIANCE value of SMU<n> for the sweep
measurement.
<n> is required to specify SMU number. Valid SMU numbers are SMU1 through
SMU6, depending on the configuration.
You use this command only if the function of the specified SMU is CONStant and
the mode is not COMMon.
At *RST, this value is:
SMU1
SMU2
SMU3
SMU4
SMU5 and SMU6
:PAGE:MEASure [:SWEep] :CONStant:SMU<n>:COMPliance compliance |
MINimum | MAXimum
Parameter
Type
compliance
numeric
compliance <newline><^END>
compliance is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:CONS:SMU1:COMP 0.1"
OUTPUT @Hp4155;":PAGE:MEAS:CONS:SMU1:COMP MAX"
OUTPUT @Hp4155;":PAGE:MEAS:CONS:SMU1:COMP?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:CONS:SMU1:COMP? MAX"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
SMU
COMPLIANCE
not defined
2 V
100 mA
100 mA
not defined
Explanation
-200 to 200 V or -1 to 1 A. The range of this value
depends on the type of SMU.
SCPI Commands
5-219