Page:ksweep:var1:Polarity - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example
Syntax
Parameter
Query response
Example
SCPI Commands

:PAGE:KSWeep:VAR1:POLarity

:PAGE:KSWeep:VAR1:POLarity
This command sets the polarity of VAR1 sweep area for knob sweep measurement.
At *RST, this value is POSitive.
:PAGE:KSWeep:VAR1:POLarity POSitive | NEGative | BIPolar
Parameter
Type
POSitive
character
NEGative
character
BIPolar
character
POS | NEG | BIP <newline><^END>
OUTPUT @Hp4155;":PAGE:KSW:VAR1:POL POS"
OUTPUT @Hp4155;":PAGE:KSW:VAR1:POL?"
ENTER @Hp4155;A$
:PAGE:KSWeep:VAR1:RANGe
This command sets the range of VAR1 for knob sweep measurement.
At *RST, this value is 1 V.
:PAGE:KSWeep:VAR1:RANGe range
Parameter
Type
range
numeric
range <newline><^END>
range is NR3 response data.
OUTPUT @Hp4155;":PAGE:KSW:VAR1:RANG 1"
OUTPUT @Hp4155;":PAGE:KSW:VAR1:RANG?"
ENTER @Hp4155;A
5-156
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
sweep in positive X direction
sweep in negative X direction
sweep in positive and negative X directions
Explanation
0.1 V to 200 V, or 1E-9 A to 1 A