:PAGE:MEASure:QSCV:VAR1:STOP

:PAGE:MEASure:QSCV:VAR1:STOP
This command sets the VAR1 sweep stop value for Quasi-static CV measurements.
This value is the upper or lower limit of the sweep output. This command also has a
query form.
At *RST, this value is set to 1.
:PAGE:MEASure:QSCV:VAR1:STOP stop | MINimum | MAXimum
:PAGE:MEASure:QSCV:VAR1:STOP?
Parameter
Type
stop
numeric
stop <newline><^END>
stop is an NR3 response data type.
In the QSCV measurement, the measurement unit executes the capacitance
measurement for the sweep steps except for the sweep start voltage and stop voltage.
At each sweep step, the capacitance measurement is executed over the voltage
range: output voltage ± capacitance measurement voltage.
The number of sweep steps is calculated from the equation (fractions below decimal
point are rounded down): |start-stop| / |step| -1. You must set the start, stop, and
step values so that the number of sweep steps value is in the range: 1 to 1001. If the
calculation result is 0, the value is automatically set to 1. Then the measurement unit
executes a one-point capacitance measurement between the start and stop values.
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:VAR1:STOP 10"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:VAR1:STOP?"
ENTER @Hp4155;A
:PAGE:MEASure:QSCV:VAR1:CSTEp
5-206
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
VAR1 sweep stop value. –200 to 200 V. The range
of this value depends on the type of VAR1 unit
being used.