Parameter
Query response
Example
Syntax
Parameter
Query response
Example
Parameter
Type
var_name
string
expr
string
var_name, expr <newline><^END>
var_name and expr are string response data, but do not contain double quote
characters at the beginning and end.
OUTPUT @Hp4155;":PAGE:DISP:ANAL:MARK:AFT 'DGM','MAX(DGM)
'"
OUTPUT @Hp4155;":PAGE:DISP:ANAL:MARK:AFT?"
ENTER @Hp4155;A$

:PAGE:DISPlay:ANALysis:MARKer:AFTer:STATe

This command selects whether the start position of search is valid.
If MARK:AFTER var_name, expr is set, the STATe is automatically set to ON.
At *RST, this value is OFF.
:PAGE:DISPlay:ANALysis:MARKer:AFTer:STATe OFF | ON | 0 | 1
Parameter
Type
OFF or 0
boolean
ON or 1
boolean
0 | 1 <newline><^END>
OUTPUT @Hp4155;":PAGE:DISP:ANAL:MARK:AFT:STAT ON"
OUTPUT @Hp4155;":PAGE:DISP:ANAL:MARK:AFT:STAT?"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:DISPlay:ANALysis:MARKer:AFTer:STATe
Explanation
data variable name
condition expression
Explanation
search entire area for the point
search only after the specified start position
SCPI Commands