:PAGE:MEASure[:SWEep]:VAR1:PCOMpliance

This command sets the power compliance value of VAR1.
If the unit type of VAR1 is VSU, this parameter is ignored.
At *RST, this value is not defined, but the PCOMpliance:STATe is OFF.
:PAGE:MEASure [ :SWEep] :VAR1:PCOMpliance pcompliance | MINimum |
MAXimum
Parameter
Type
pcompliance
numeric
If SMU for VAR1 unit is set to VPULSE or IPULSE mode, you cannot set power
compliance for VAR1 unit.
pcompliance <newline><^END>
pcompliance is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:PCOM 0.5"
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:PCOM MAX"
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:PCOM?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:PCOM? MAX"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:MEASure[:SWEep]:VAR1:PCOMpliance
Explanation
1E-3 to 20. The range of this value depends on the
unit type of VAR1.
SCPI Commands
5-227