Query response
Example
Syntax
Example
Syntax
Parameter
Query response
Example
hue, sat, lum <newline><^END>
The data type of hue, sat, and lum are NR3 response data.
OUTPUT @Hp4155;":DISP:CMAP:COLor1:HSL 0.5,0.5,0.5"
OUTPUT @Hp4155;":DISP:CMAP:COLor1:HSL?"
ENTER @Hp4155;A,B,C

:DISPlay:CMAP:DEFault

This command sets the color map to the instrument's default values for all colors.
This command does not have query form.
:DISP:CMAP:DEFault
OUTPUT @Hp4155;":DISP:CMAP:DEF"

:DISPlay:CONTrast

The 4155C/4156C does not do anything by this command. This command is only
for keeping the compatibility with the 4155A/4156A.
AT *RST, this value is 0.608.
:DISPlay:CONTrast contrast | MINimum | MAXimum
Parameter
Type
contrast
numeric
contrast <newline><^END>
The data type of contrast is NR3 response data.
OUTPUT @Hp4155;":DISP:CONT 0.5"
OUTPUT @Hp4155;":DISP:CONT?"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
0 to 1. 0 means no difference between data and
background, and 1 means maximum contrast.
SCPI Commands
:DISPlay:CMAP:DEFault
5-27