Status:questionable:ptransition - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Semantics
Example

:STATus:QUEStionable:PTRansition

This command sets the positive transition filter of the Questionable Status
"CONDITION" register.
:STATus:QUEStionable:PTRansition register
Parameter
Type
register
numeric or
non-decim
al numeric
register <newline><^END>
register is a decimal integer value, which is the sum of the binary-weighted values,
in NR1 response data format.
If a bit in the positive transition filter is set to 1, then a 0 to 1 transition in the
corresponding bit of the Questionable Status "CONDITION" register causes a 1 to
be written in the corresponding "EVENT" register.
OUTPUT @Hp4155;":STAT:QUES:PTR 16384"
OUTPUT @Hp4155;":STAT:QUES:PTR #H4000"
OUTPUT @Hp4155;":STAT:QUES:PTR #Q40000"
OUTPUT @Hp4155;":STAT:QUES:PTR #B100000000000000"
OUTPUT @Hp4155;":STAT:QUES:PTR?"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:STATus:QUEStionable:PTRansition
Explanation
decimal integer, hexadecimal, octal, or binary value
that is the sum of the binary-weighted values for the
desired bits.
SCPI Commands
5-299

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4156c

Table of Contents