Error -300
Error -310
Error -311
Error -313
Error -315
Error -330
Error -350

Device-Dependent Error

-3XX errors indicate that the 4155C/4156C has detected an error that is not a
command error, a query error, or an execution error; some device operations did not
properly complete, possibly due to an abnormal hardware or firmware condition.
These codes are also used for self-test response errors.
Device-specific error
Generic device-dependent error for the 4155C/4156C that cannot be determined
more specifically.
System error
Some error, termed "system error" by the 4155C/4156C, has occurred.
Memory error
An error was detected in the 4155C/4156C's memory.
Calibration memory lost
Nonvolatile data saved by the *CAL? command has been lost.
Configuration memory lost
Nonvolatile configuration data saved by the 4155C/4156C has been lost.
Self-test failed
The self-test has failed.
Queue overflow
This code is entered into the queue instead of the code that caused the error. This
code indicates that there is no room in the queue and an error occurred but was not
recorded.