Syntax
Example
Syntax
Parameter
Query response
Example

:PAGE:GLISt[:GRAPhics]:AANalysis

This command redisplays the marker, cursors, and lines at the same positions as
after auto-analysis was performed when the measurement finished. Auto-analysis is
determined by DISPLAY: ANALYSIS SETUP, which corresponds to the
:PAGE:DISPLAY:ANALysis subsystem.
This command does not have query form.
:PAGE:GLISt [ :GRAPhics] :AANalysis
OUTPUT @Hp4155;":PAGE:GLIS:AAN"
:PAGE:GLISt[:GRAPhics]:ASELect
This command selects whether the Y axis is Y1 or Y2.
At *RST, this value is Y1.
:PAGE:GLISt [ :GRAPhics] :ASELect Y1 | Y2
Parameter
Type
Y1
character
Y2
character
Y1 | Y2 <newline><^END>
OUTPUT @Hp4155;":PAGE:GLIS:ASEL Y1"
OUTPUT @Hp4155;":PAGE:GLIS:ASEL?"
ENTER @Hp4155;A$
:PAGE:GLISt[:GRAPhics]:CURSor:DIRect:X|:Y1|:Y2
This command moves the cursor to the point that has the specified X, Y1, or Y2
value.
For the query response, the returned data is the X value of the present cursor
position.
Variable name must be assigned to the specified axis by using
:PAGE:DISP:GRAP:X|:Y1|:Y2:NAME command.
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:GLISt[:GRAPhics]:AANalysis
Explanation
y axis is Y1
y axis is Y2
SCPI Commands