Syntax
Parameter
Query response
Semantics
Example
Syntax
Example
:STATus:OPERation:PTRansition
This command sets the positive transition filter for the Operation Status
"CONDITION" register.
:STATus:OPERation:PTRansition register
Parameter
Type
register
numeric or
non-decim
al numeric
register <newline><^END>
register is a decimal integer value, which is the sum of the binary-weighted values,
in NR1 response data format.
If a bit in the positive transition filter is set to 1, then a 0 to 1 transition in the
corresponding bit of the Operation Status "CONDITION" register causes a 1 to be
written in the corresponding "EVENT" register.
OUTPUT @Hp4155;":STAT:OPER:PTR 16384"
OUTPUT @Hp4155;":STAT:OPER:PTR #H4000"
OUTPUT @Hp4155;":STAT:OPER:PTR #Q40000"
OUTPUT @Hp4155;":STAT:OPER:PTR #B100000000000000"
OUTPUT @Hp4155;":STAT:OPER:PTR?"
ENTER @Hp4155;A
:STATus:PRESet
This command presets the event enable (MASK) register and TRANSITION
FILTER registers of the Operation Status, Questionable Status, and Emergency
Status registers.
This command does not have query form.
:STATus:PRESet
OUTPUT @Hp4155;":STAT:PRES"
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:STATus:OPERation:PTRansition
Explanation
decimal integer, hexadecimal, octal, or binary value
that is the sum of the binary-weighted values for the
desired bits.
SCPI Commands
5-295