Page:measure:qscv:var1:Compliance - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example

:PAGE:MEASure:QSCV:VAR1:COMPliance

This command sets the compliance value of the VAR1 channel for Quasi-static CV
measurements. This command also has a query form.
:PAGE:MEASure:QSCV:VAR1:COMPliance compliance | MINimum | MAXimum
:PAGE:MEASure:QSCV:VAR1:COMPliance?
Parameter
Type
compliance
numeric
compliance <newline><^END>
compliance is an NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:VAR1:COMP 0.1"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:VAR1:COMP?"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:MEASure:QSCV:VAR1:COMPliance
Explanation
VAR1 compliance value. –1 to 1 A. The range of
this value depends on the type of VAR1 unit being
used.
SCPI Commands
5-201

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4156c