Syntax
Parameter
Query response
Example
SCPI Commands
:PAGE:STRess:SETup:SSTop
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:WIDTH? MIN"
ENTER @Hp4155;A
:PAGE:STRess:SETup:SSTop
This command sets the stress stop mode for an abnormal status.
The abnormal statuses that can be detected are as follows:
•
SMU reaches its compliance setting.
•
Current of VSU exceeds ±100 mA.
•
SMU or VSU oscillates.
•
A/D converter overflow occurs.
•
Average current of PGU exceeds ±100 mA.
At *RST, this value is OFF.
:PAGE:STRess:SETup:SSTop ABNormal | COMPliance | OFF
Parameter
Type
ABNormal
character
COMPliance
character
OFF
character
ABN | COMP | OFF <newline><^END>
OUTPUT @Hp4155;":PAGE:STR:SET:SST ABN"
OUTPUT @Hp4155;":PAGE:STR:SET:SST?"
ENTER @Hp4155;A$
5-278
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
Stop stress if abnormal status is detected.
Stops stress only if some SMU reaches its
compliance setting.
Continue stress even if abnormal status is
detected.