Page:measure:osequence:trigger:polarity - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example
Syntax
Parameter
Query response
Example
SCPI Commands

:PAGE:MEASure:OSEQuence:TRIGger:POLarity

:PAGE:MEASure:OSEQuence:TRIGger:POLarity
This command sets the POLARITY of trigger signal.
At *RST, this value is POSitive.
:PAGE:MEASure:OSEQuence:TRIGger:POLarity POSitive | NEGative
Parameter
Type
POSitive
character
NEGative
character
POS | NEG <newline><^END>
OUTPUT @Hp4155;":PAGE:MEAS:OSEQ:TRIG:POL POS"
OUTPUT @Hp4155;":PAGE:MEAS:OSEQ:TRIG:POL?"
ENTER @Hp4155;A$
:PAGE:MEASure:OSEQuence:TRIGger[:STATe]
This command controls whether the trigger function is enabled.
At *RST, this value is OFF.
:PAGE:MEASure:OSEQuence:TRIGger [ :STATe] OFF | ON | 0 | 1
Parameter
Type
OFF or 0
boolean
ON or 1
boolean
0 | 1 <newline><^END>
OUTPUT @Hp4155;":PAGE:MEAS:OSEQ:TRIG ON"
OUTPUT @Hp4155;":PAGE:MEAS:OSEQ:TRIG?"
ENTER @Hp4155;A
5-176
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
polarity is positive
polarity is negative
Explanation
trigger function is disabled
trigger function is enabled

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4156c

Table of Contents