Page:ksweep:var1:Stime - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example

:PAGE:KSWeep:VAR1:STIMe

This command sets the step time (length of each step) for knob sweep measurement.
At *RST, this value is 0.5 ms.
:PAGE:KSWeep:VAR1:STIMe step_time | MINimum | MAXimum
Parameter
Type
step_time
numeric
step_time <newline><^END>
step_time is NR3 response data type.
OUTPUT @Hp4155;":PAGE:KSW:VAR1:STIM 0.005"
OUTPUT @Hp4155;":PAGE:KSW:VAR1:STIM MAX"
OUTPUT @Hp4155;":PAGE:KSW:VAR1:STIM?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:KSW:VAR1:STIM? MAX"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:KSWeep:VAR1:STIMe
Explanation
0.5E-3 to 100E-3 s (Resolution: 100E-6)
SCPI Commands
5-157