Page:measure[:Sweep]:Constant:vsu<N>[:Source] - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example
:PAGE:MEASure[:SWEep]:CONStant:VSU<n>[:SOURce]
:PAGE:MEASure[:SWEep]:CONStant:VSU<n>[:SOU
Rce]
This command sets the constant SOURCE value of VSU<n> for the sweep
measurement.
<n> is required to specify VSU number. Valid VSU numbers are VSU1 through
VSU2.
You use this command only if the function of the specified VSU is CONStant.
At *RST, this value is 0V.
:PAGE:MEASure [ :SWEep] :CONStant:VSU<n>[ :SOURce] source | MINimum |
MAXimum
Parameter
Type
source
numeric
source <newline><^END>
source is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:CONS:VSU1 10"
OUTPUT @Hp4155;":PAGE:MEAS:CONS:VSU1 MAX"
OUTPUT @Hp4155;":PAGE:MEAS:CONS:VSU1?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:CONS:VSU1? MAX"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
-20 to 20 V
SCPI Commands
5-221