Page:stress:setup:pulse:pgu<N>:Impedance - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Query response
Example
Syntax
Parameter
Query response
Example
delay_time <newline><^END>
delay_time is NR3 response data type.
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:DEL 1"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:DEL MAX"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:DEL?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:DEL? MAX"
ENTER @Hp4155;A
:PAGE:STRess:SETup:PULSe:PGU<n>:IMPedance
This command sets the output IMPEDANCE of PGU<n> for stress.
<n> is required to specify PGU number. Valid PGU numbers are PGU1 through
PGU2.
The mode of the specified PGU must be VPULse.
At *RST, this value is LOW.
:PAGE:STRess:SETup:PULSe:PGU<n>:IMPedance LOW | R50
Parameter
Type
LOW
character
R50
character
LOW | R50 <newline><^END>
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:IMP LOW"
OUTPUT @Hp4155;":PAGE:STR:SET:PULS:PGU1:IMP?"
ENTER @Hp4155;A$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:STRess:SETup:PULSe:PGU<n>:IMPedance
low impedance
50 W
SCPI Commands
Explanation
5-273