Syntax
Parameter
Query response
Example
Syntax
Query response
Example
:PAGE:CHANnels:MATrix:GPIB:ADDRess
This command sets the GPIB address of the Agilent E5250A Low Leakage Switch
Mainframe to be controlled from the 4155C/4156C. This command also has a query
form.
At *RST, this value retains its previous value.
:PAGE:CHANnels:MATrix:GPIB:ADDRess address
:PAGE:CHANnels:MATrix:GPIB:ADDRess?
Parameter
Type
address
numeric
address <newline><^END>
address is NR1 response data type.
OUTPUT @Hp4155;":PAGE:CHAN:MAT:GPIB:ADDR 10"
OUTPUT @Hp4155;":PAGE:CHAN:MAT:GPIB:ADDR?"
ENTER @Hp4155;A
:PAGE:CHANnels:UFUNction:CATalog?
This query command returns the number of and a list of currently defined functions.
This command has query form only.
:PAGE:CHANnels:UFUNction:CATalog?
num_function , function_list <newline><^END>
num_function is a numeric value, which is the number of user defined functions.
The data type is NR1 response data.
function_list is a list of the user function names separated by commas. The data type
is string response data, but does not contain double quote characters at the beginning
and end of the string.
OUTPUT @Hp4155;":PAGE:CHAN:UFUN:CAT?"
ENTER @Hp4155;A,B$
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:CHANnels:MATrix:GPIB:ADDRess
Explanation
GPIB address of the Agilent E5250A. 0 to 30.
Initial setting: 22.
SCPI Commands
5-101