Syntax
Parameter
Query response
Example
Syntax
Example

:PAGE:GLISt[:GRAPhics]:DISPlay:DVARiables

This command selects whether the variables specified in the DATA VARIABLES
fields of the DISPLAY: DISPLAY SETUP will be displayed on the graph or list.
At *RST, this value is OFF.
:PAGE:GLISt [ :GRAPhics] :DISPlay:DVARiables OFF | ON | 0 | 1
Parameter
Type
OFF or 0
boolean
ON or 1
boolean
0 | 1 <newline><^END>
OUTPUT @Hp4155;":PAGE:GLIS:DISP:DVAR ON"
OUTPUT @Hp4155;":PAGE:GLIS:DISP:DVAR?"
ENTER @Hp4155;A
:PAGE:GLISt[:GRAPhics]:DISPlay:OVERlay:STOVer
lay
This command forces the scale of the present measurement data to that of the
overlay plane.
This command does not have query form.
:PAGE:GLISt [ :GRAPhics] :DISPlay:OVERlay:STOVerlay
OUTPUT @Hp4155;":PAGE:GLIS:DISP:OVER:STOV"
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:GLISt[:GRAPhics]:DISPlay:DVARiables
Explanation
variable list is off
variable list is on
SCPI Commands
5-135