Syntax
Parameter
Query response
Example
Syntax
Parameter
Query response
Example
SCPI Commands

:PAGE:KSWeep:DISPlay:REGion:X

:PAGE:KSWeep:DISPlay:REGion:X
This command sets the display region of X axis for knob sweep.
At *RST, this value is POSitive.
:PAGE:KSWeep:DISPlay:REGion:X POSitive | NEGative | BIPolar
Parameter
Type
POSitive
character
NEGative
character
BIPolar
character
POS | NEG | BIP <newline><^END>
OUTPUT @Hp4155;":PAGE:KSW:DISP:REG:X POS"
OUTPUT @Hp4155;":PAGE:KSW:DISP:REG:X?"
ENTER @Hp4155;A$
:PAGE:KSWeep:DISPlay:REGion:Y
This command sets the display region of Y axis for knob sweep.
At *RST, this value is POSitive.
:PAGE:KSWeep:DISPlay:REGion:Y POSitive | NEGative | BIPolar
Parameter
Type
POSitive
character
NEGative
character
BIPolar
character
POS | NEG | BIP <newline><^END>
OUTPUT @Hp4155;":PAGE:KSW:DISP:REG:Y POS"
OUTPUT @Hp4155;":PAGE:KSW:DISP:REG:Y?"
ENTER @Hp4155;A$
5-152
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
display positive X region
display negative X region
display positive and negative X region
Explanation
display positive Y region
display negative Y region
display positive and negative Y region