Syntax
Parameter
Query response
Example
SCPI Commands
:PAGE:CHANnels[:CDEFinition]:VMU<n>:DCHarge
:PAGE:CHANnels[:CDEFinition]:VMU<n>:DCHarge
This command connects, or disconnects, the VMU input discharge resistor used to
prevent the VMU inputs from becoming charged when the inputs are opened. This
command also has a query form.
At *RST, this value is ON.
:PAGE:CHANnels [:CDEFinition] :VMU<n>:DCHarge OFF | ON | 0 | 1
:PAGE:CHANnels [:CDEFinition] :VMU<n>:DCHarge?
<n> must be 1 or 2.
Parameter
Type
OFF or 0
Boolean
ON or 1
Boolean
When an auto-calibration is executed, the setting is not changed.
0 | 1 <newline><^END>
OUTPUT @Hp4155;":PAGE:CHAN:VMU1:DCH ON"
The command above specifies VMU1, but it connects the discharge resistor to both
the VMU1 and VMU2 inputs.
OUTPUT @Hp4155;":PAGE:CHAN:VMU1:DCH?"
ENTER @Hp4155;A
5-94
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
Disconnects the discharge resistors from the VMU
inputs.
Connects the discharge resistors to the VMU inputs,
and automatically breaks the connection only when
in the measurement state.