Page:measure:qscv:delay - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Parameter
Query response
Example

:PAGE:MEASure:QSCV:DELay

This command sets the DELAY TIME for Quasi-static CV measurements. The
delay time is the time from the start of each sweep step to the start of the
measurement. This command also has a query form.
At *RST, this value is set to 0.
:PAGE:MEASure:QSCV:DELay delay_time
:PAGE:MEASure:QSCV:DELay?
Parameter
Type
delay_time
numeric
delay_time <newline><^END>
delay_time is an NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:DEL 1.5"
OUTPUT @Hp4155;":PAGE:MEAS:QSCV:DEL?"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:MEASure:QSCV:DELay
Explanation
0 to 65.535 (s). Resolution: 100 ms.