Syntax
Parameter
Query response
Remarks
Example
:DISPlay[:WINDow][:STATe]
This command controls whether the instrument screen is updated or not.
AT *RST, this value is ON.
:DISPlay[ :WINDow] [ :STATe] OFF | ON | 0 | 1
Parameter
Type
OFF or 0
boolean
ON or 1
boolean
0 | 1 <newline><^END>
The :DISP OFF (or 0) command is not available when the 4155C/4156C screen
displays the following page:
•
SYSTEM: FILER
•
SYSTEM: MISCELLANEOUS
•
SYSTEM: CONFIGURATION
•
SYSTEM: SELF-CALIBRATION/DIAGNOSTICS
•
SYSTEM: PRINT/PLOT SETUP
•
SYSTEM: COLOR SETUP
•
KNOB SWEEP
To enter the following command, the 4155C/4156C screen must be set to the update
enable state.
•
:PAGE:KSWeep command group
•
:PAGE:SCONtrol:KSWeep[:STARt] command
•
:DIAGnostic:TEST[:EXECute] test_no command (test_no: 201 to 413)
OUTPUT @Hp4155;":DISP ON"
OUTPUT @Hp4155;":DISP?"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:DISPlay[:WINDow][:STATe]
Explanation
screen is not updated
screen is updated (update enable state)
SCPI Commands