:PAGE:MEASure:OSEQuence[:MENU]

:PAGE:MEASure:OSEQuence[:MENU]
This command changes the present display page to MEASURE: OUTPUT
SEQUENCE page. This command does not have query form.
:PAGE:MEASure:OSEQuence [ :MENU]
OUTPUT @Hp4155;":PAGE:MEAS:OSEQ"
:PAGE:MEASure:OSEQuence:MODE
This command selects the output sequence mode for sampling measurement.
You use this command only if the measurement mode is sampling.
At *RST, this value is SIMultaneous for sampling measurement.
:PAGE:MEASure:OSEQuence:MODE SEQuential | SIMultaneous
Parameter
Type
SEQuential
character
SIMultaneous
character
SEQ | SIM <newline><^END>
OUTPUT @Hp4155;":PAGE:MEAS:OSEQ:MODE SIM"
OUTPUT @Hp4155;":PAGE:MEAS:OSEQ:MODE?"
ENTER @Hp4155;A$
:PAGE:MEASure:OSEQuence:OSEQuence
This command sets the OUTPUT SEQUENCE.
The source channels output in the order specified by the OUTPUT SEQUENCE.
Unspecified units output in default order after the specified units.
At *RST, the output sequence is SMU1-SMU2-SMU3-SMU4-SMU5-SMU6-
VSU1-VSU2-PGU1-PGU2.
:PAGE:MEASure:OSEQuence:OSEQuence unit { , unit }| DEFault