Page:measure:msetup:itime:long:aperture? - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Syntax
Query response
Example
Syntax
Parameter
Query response
Example

:PAGE:MEASure:MSETup:ITIMe:LONG:APERture?

This command returns the LONG integration time in terms of the APERTURE,
which means time.
The APERture value is related to NPLCycles by the following equation:
APERture = NPLCycles / selected line frequency
This command has query form only.
At *RST, this value is 16/selected line frequency.
:PAGE:MEASure:MSETup:ITIMe:LONG:APERture?
aperture <newline><^END>
aperture is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:LONG:APER?"
ENTER @Hp4155;A
:PAGE:MEASure:MSETup:ITIMe:LONG[:NPLCycle]
This command sets the LONG integration time in terms of the Number of Power
Line Cycles. At *RST, this value is 16.
:PAGE:MEASure:MSETup:ITIMe:LONG [ :NPLCycle] nplcycle | MINimum |
MAXimum
Parameter
Type
nplcycle
numeric
nplcycle <newline><^END>
nplcycle is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:LONG 16"
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:LONG MAX"
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:LONG?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:MSET:ITIM:LONG? MAX"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1
:PAGE:MEASure:MSETup:ITIMe:LONG:APERture?
2 to 100
SCPI Commands
Explanation
5-165

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4156c

Table of Contents