Syntax
Parameter
Query response
Example
Syntax
Parameter
SCPI Commands

:PAGE:MEASure[:SWEep]:VAR1:COMPliance

:PAGE:MEASure[:SWEep]:VAR1:COMPliance
This command sets the COMPLIANCE value of VAR1.
If the unit type of VAR1 is VSU, this parameter is ignored.
At *RST, this value is 100 mA.
:PAGE:MEASure [ :SWEep] :VAR1:COMPliance compliance | MINimum |
MAXimum
Parameter
Type
compliance
numeric
compliance <newline><^END>
compliance is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:COMP 0.1"
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:COMP MAX"
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:COMP?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:VAR1:COMP? MAX"
ENTER @Hp4155;A
:PAGE:MEASure[:SWEep]:VAR1:MODE
This command sets the SWEEP MODE of VAR1 for normal (not knob) sweep
measurement. At *RST, this value is SINGle.
:PAGE:MEASure [ :SWEep] :VAR1:MODE SINGle | DOUBle
Parameter
Type
SINGle
character
DOUBle
character
5-226
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
-200 V to 200V or -1 A to 1 A. The range of this
value depends on the unit type of VAR1.
Explanation
single stair sweep
double stair sweep