Page:measure[:Sweep]:Var2:Pcompliance:state - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Parameter
Query response
Example
Syntax
Parameter
Query response
Example
Parameter
Type
pcompliance
numeric
pcompliance <newline><^END>
pcompliance is NR3 response data type.
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:PCOM 0.5"
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:PCOM MAX"
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:PCOM?"
ENTER @Hp4155;A
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:PCOM? MAX"
ENTER @Hp4155;A
:PAGE:MEASure[:SWEep]:VAR2:PCOMpliance:STA
Te
This command sets the power compliance of VAR2 to disable.
You use this command only if there is an SMU whose function (FCTN) is VAR2.
If the unit type of VAR2 is VSU, this parameter is ignored.
If power compliance value is set by PCOMpliance pcompliance, the query returns
1.
At *RST, this value is OFF.
:PAGE:MEASure [ :SWEep] :VAR2:PCOMpliance:STATe OFF | 0
Parameter
Type
OFF or 0
boolean
0 | 1 <newline><^END>
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:PCOM:STATE OFF"
OUTPUT @Hp4155;":PAGE:MEAS:VAR2:PCOM:STATE?"
ENTER @Hp4155;A
Agilent 4155C/4156C SCPI Command Reference, Edition 1

:PAGE:MEASure[:SWEep]:VAR2:PCOMpliance:STATe

1E-3 to 20. The range of this value depends on
SMU type of VAR2.
power compliance is disabled
SCPI Commands
Explanation
Explanation
5-233