Syntax
Parameter
Query response
Example
See also
Syntax
SCPI Commands

:HCOPy:LINDex

:HCOPy:LINDex
This command specifies the range of measurement results to output by using index
numbers.
You use this command only if the current display page is GRAPH/LIST: LIST.
This command also changes the value of :MMEMory:STORe:SSHeet:LINDex.
At *RST, this value is set to (1,MAX).
This command does not comply with SCPI.
:HCOPy:LINDex first | MINimum | MAXimum, last | MINimum | MAXimum
Parameter
Type
first
numeric
last
numeric
first,last <newline><^END>
The data type of first and last are NR1 response data.
OUTPUT @Hp4155;":HCOP:LIND 1,MAX"
OUTPUT @Hp4155;":HCOP:LIND?"
ENTER @Hp4155;A,B
:MMEMory:STORe:SSHeet:LINDex

:HCOPy:OPAGe

This command sets the range of the setup data to be output.
At *RST, this value is set to CURRent.
:HCOPy:OPAGe CURRent | GROup | ALL
5-52
Agilent 4155C/4156C SCPI Command Reference, Edition 1
Explanation
first index number to output
last index number to output